Abstract:
      用不同宽度袖带测定一组儿童血压,证实了袖带宽度可明显影响血压测得值的精确度,并由此推导出不适宜袖带测压值的校正公式。在另两组儿童血压测定中检验了该校正公式的有效性及可行性。血压校正公式的提出,使实施美国心脏学会建议的臂中点周径百分之四十为适宜袖宽的标准成为可能。血压校正公式应用简便,可在儿童医疗预防、卫生保健、特别是各种血压调查、科研协作中发挥积极作用。
English Abstract:
      The blood pressures of a group of children was obtained with sphygmomanomiter cuffs of varying widths. It was suggested that the cuff width can be effect for accuracy of blood pressure determination,then a correct formula of blood pressure determination using an inadequate cuffs was calculated.
The efficacy and feasibility of the formula was tested in another two groups of children.The production of the formula made a possibility for practicing the proposal of American Heart Association in which the width of an adequate cuff was believed to be 40% of the circumference in middle arms.The formula was simple and convenient and should be used in childhood prophylactic medicine, healthy care especially in popularized measurement of children blood pressure and scientific cooperative investigation.