Abstract
吴星,周诚,黄维金,祁自柏,梁争论,李河民,庄辉.对四种国产乙型肝炎酶联免疫试剂盒的灵敏度和特异度分析[J].Chinese journal of Epidemiology,2008,29(9):915-918
对四种国产乙型肝炎酶联免疫试剂盒的灵敏度和特异度分析
Sensitivity and specificity of 4 domestic ELISA kits for detection of hepatitis B virus markers
Received:January 05, 2008  
DOI:
KeyWord: 乙型肝炎病毒|诊断试剂|酶联免疫吸附试验
English Key Word: Hepatitis B virus|Diagnostic kits|ELISA
FundProject:国家"十五"科技攻关计划资助项目(2004BA718B02)
Author NameAffiliationE-mail
WU Xing National Institute for the Control of Pharmaceutical and Biological Products, Beijing 100050, China  
ZHOU Cheng National Institute for the Control of Pharmaceutical and Biological Products, Beijing 100050, China  
HUANG Wei-jin National Institute for the Control of Pharmaceutical and Biological Products, Beijing 100050, China  
QI Zibai National Institute for the Control of Pharmaceutical and Biological Products, Beijing 100050, China  
LIANG Zheng-lun National Institute for the Control of Pharmaceutical and Biological Products, Beijing 100050, China  
LI He-min National Institute for the Control of Pharmaceutical and Biological Products, Beijing 100050, China  
ZHUANG Hui 北京大学医学部 zhuangbmu@126.com 
Hits: 3739
Download times: 1391
Abstract:
      目的 比较4种国产乙型肝炎(乙肝)ELISA试剂盒的灵敏度和特异度。方法 取慢性乙肝患者和不合格献血员血液样本594份,应用4种国产乙肝ELISA试剂盒检测HBsAg、抗-HBs、HBeAg、抗-HBe和抗-HBc,并与1种国外乙肝化学发光试剂盒(Abbott公司Architect)比较,对结果不一致的样本再用各试剂盒做双孔重复检测,对HBsAg 临界值附近的样本用Abbott HBsAg确证试剂盒(Architect HBsAg confirm)验证;以Abbott试剂盒(Architect)的检测结果为标准,计算4种国产ELISA试剂盒的灵敏度、特异度、总符合率和约登指数.然后用上述试剂盒榆测HBV 5项标志物的灵敏度参比品,比较不同试剂盒的灵敏度。结果与Abbott公司Architect试剂盒相比,国产HBsAgELISA试剂盒的灵敏度低4~10倍;抗-HBs、HBeag、抗.HBe和抗-HBc ELISA试剂盒的灵敏度低4~16倍;且有假阳性.HBsAg、抗-HBs、HBeAg、抗-HBe和抗-HBc的总符合率分别为96.46%~98.15%、94.28%~98.15%、98.15%~99.49%、90.07%~96.30%、92.09%~96.80%。结论国产乙肝ELISA试剂盒的灵敏度和特异度有待进一步提高。
English Abstract:
      Objective To compare and analyze the sensitivity,specificity of 4 domestic ELISA kits for detection of hepatitis B virus (HBV) markers (HBsAg,anti-HBs,HBeAg,anti-HBe,and anti-HBc).Methods Five hundred and ninety four serum samples collected from patients with chronic hepatitis B and abnormal blood donors were detected for HBV markers and by 4 domestic ELISA kits.Samples with conflicting results by different diagnostic kits were retested.Samples with the HBsAg values close to the cut-off point were detected by Abbott HBsAg confirmation kit (Architect HBsAg confirm).Sensitivity of the kits was determined,using the national sensitivity reference panels for HBsAg,anti-HBs,HBeAg,anti-HBe and anti-HBc.Results The rates of sensitivity on 4 domestic kits for detection of HBsAg were 4 to 10 times lower,and on the 4 domestic kits for detection of anti-HBs,HBeAg,anti-HBc and anti-HBc were 4 to 16 times lower,as compared to Abbott Architect kits.In addition,the domestic HBV ELISA kits had some false positive results.The total coincidence rates of HBsAg,anti-HBs,HBeAg,anti-HBe,anti-HBc were 96.46%-98.15%,94.28%-98.15%,98.15%-99.49%,90.07%-96.30%,92.09%-96.80%,respectively.Conclusion Both sensitivity and specificity of the domestically produced HBV ELISA kits should be improved.
View Fulltext   Html FullText     View/Add Comment  Download reader